• Chinese Optics Letters
  • Vol. 24, Issue 10, 103001 (2026)
Yingchao Xie1, Yuqing Li2, Hongjin Liu2, Guishi Wang2, Gaopeng Lu1, Kun Liu2、*, and Chunyan Sun3、**
Author Affiliations
  • 1School of Earth and Space Sciences, University of Science and Technology of China, Hefei 230026, China
  • 2Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei 230031, China
  • 3School of Physics and Astronomy, Anqing Normal University, Anqing 246133, China
  • *Corresponding author: liukun@aiofm.ac.cn, scy05906@mail.ahnu.edu.cn
  • show less
    DOI: 10.3788/COL202624.103001 Cite this Article Set citation alerts
    Yingchao Xie, Yuqing Li, Hongjin Liu, Guishi Wang, Gaopeng Lu, Kun Liu, Chunyan Sun, "Water-vapor-enhanced off-beam QEPAS sensor for in-situ monitoring of dissolved CO2," Chin. Opt. Lett. 24, 103001 (2026) Copy Citation Text show less

    Abstract

    Sensitive dissolved CO2 sensors are essential for aquatic studies, but existing devices often trade sensitivity for portability. Here we present a portable sensor integrating off-beam quartz-enhanced photoacoustic spectroscopy (QEPAS) with vacuum membrane degassing. A key innovation is using water vapor as a signal amplifier. Precise temperature control maintains a saturated gas stream, boosting the photoacoustic signal by 1.5 times. An endpoint baseline correction algorithm effectively suppresses instrumental drift. The system achieves a 4.13 ppm detection limit with excellent linearity. Field stability was confirmed through in-situ monitoring in Lake Chaohu. This versatile system is suitable for environmental monitoring and carbon storage leakage detection.
    © The Authors. Published by Chinese Laser Press under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.

    1. Introduction

    Carbon dioxide (CO2) dissolved in aquatic systems constitutes a critical and dynamic reservoir within the global carbon cycle. Accurate quantification is essential for environmental studies and for resolving fundamental geochemical and geophysical processes[1,2]. In natural waters, constraining CO2 fluxes from lakes and rivers to the atmosphere is necessary for closing the global carbon budget[3,4]. Dissolved CO2 also functions as a geochemical tracer; anomalous degassing in groundwaters and geothermal fluids is often associated with deep-seated processes such as magma migration and fault activity, serving as a potential precursor to seismic events[5,6]. As a primary agent in water–rock interactions, dissolved CO2 regulates the dissolution kinetics of carbonate rocks, drives karst formation, and influences the chemical composition of subsurface fluids[7,8]. In the context of carbon capture and storage (CCS), detecting minute increases in dissolved CO2 in overlying aquifers is required for early warning of reservoir leakage[9].

    Despite the broad scientific relevance of high-precision, high-spatiotemporal-resolution data acquisition, detecting dissolved CO2 using existing methods is analytically challenging. The canonical discrete method, headspace equilibrium followed by gas chromatography, is unsuitable for capturing transient signals or fine spatial heterogeneity due to its low temporal resolution and labor-intensive protocol[10,11]. Commercial field sensors are available but often exhibit insufficient sensitivity, long-term drift, or limited selectivity when applied to complex environmental matrices[12,13]. Optical sensing offers a viable route forward. Tunable diode laser absorption spectroscopy (TDLAS) provides high sensitivity but typically uses multi-pass cells susceptible to misalignment under field-induced vibrations, compromising measurement stability and reliability[14,15]. Cavity ring-down spectroscopy (CRDS) achieves exceptionally low detection limits through extended optical path length but is limited by high cost, operational complexity, and susceptibility to environmental perturbations, restricting deployment in extensive field campaigns or remote areas[16,17]. Fluorescence-based sensors[18] and electrochemical systems[19] are highly portable but frequently suffer from signal drift, cross-sensitivity, and inadequate detection limits for resolving subtle geochemical gradients.

    Quartz-enhanced photoacoustic spectroscopy (QEPAS) has emerged as a competitive technology for trace gas sensing due to its compact design, inherent immunity to beam misalignment, and high-Q resonant detection[20,21]. A key advantage of QEPAS is its minimal sample volume requirement. The dead volume of the acoustic detection module (tuning fork and micro-resonators) is typically a few cubic centimeters, compared with the hundreds required to purge and stabilize TDLAS multi-pass cells or CRDS cavities, leading to longer system response times and higher carrier gas demands[22,23]. In dissolved gas analysis, efficient and stable gas extraction is essential for optimal optical detection. This study employs a polydimethylsiloxane (PDMS) hollow fiber membrane as the degassing element. This material provides high permeability and selectivity for non-polar gases such as CO2, enabling rapid and continuous separation of dissolved gases from the aqueous to the gas phase[24].

    The conventional on-beam QEPAS configuration, which requires precise focusing of the laser beam through the narrow gap (<500  μm) between the tuning fork prongs, achieves high sensitivity in laboratory settings. This precise optical alignment increases vulnerability to performance degradation caused by temperature fluctuations and mechanical vibrations during field deployment, limiting instrumental ruggedness and long-term stability[25]. The narrow resonance profile of the high-Q quartz tuning fork, while beneficial for noise rejection, also makes signal amplitude highly sensitive to drifts in the laser wavelength modulation frequency. If uncorrected, such drifts can cause significant signal loss and measurement instability over time, particularly under variable environmental conditions in field applications[26,27]. Active frequency tracking techniques, such as using a phase-locked loop (PLL) to lock the modulation frequency to the fork’s resonant frequency in real time or periodically sweeping the modulation frequency to locate the maximum response, are commonly implemented to maintain optimal and stable operation[28,29]. The off-beam QEPAS (OB-QEPAS) configuration was developed to overcome the optical alignment limitation[30]. By directing the laser beam adjacent to the tuning fork, alignment tolerances are substantially relaxed and mechanical robustness is improved, making the sensor architecture suitable for long-term deployment in demanding field environments, including mobile platforms and seismically active regions[31,32].

    The influence of water vapor on QEPAS performance depends on the spectral line and can be managed strategically. Spectral line selection using the HITRAN database[33,34] confirms that direct absorption interference from H2O is negligible at the 2.004 µm CO2 absorption line used in this study. Water vapor molecules act as efficient vibrational energy transfer partners for excited CO2, amplifying the photoacoustic signal through resonant vibrational-to-translational (V–T) energy relaxation[33,34]. Instead of treating water vapor as an interferent requiring removal, which reduces signal strength, this work harnesses the “water vapor enhancement effect” as a stable signal amplification mechanism. Precise thermal control of the gas stream exiting the membrane equilibrator[35] is used to actively maintain consistently high water vapor saturation. This approach stabilizes the measurement matrix and provides substantial and consistent enhancement of the CO2 photoacoustic signal, improving in-situ detection sensitivity[36].

    Combining the rugged OB-QEPAS architecture with efficient sample pretreatment via PDMS membrane degassing, controlled humidity-based signal enhancement, advanced signal processing for drift and noise suppression, and robust frequency stabilization, this study develops a portable analyzer for dissolved CO2. Field capability was validated through a field validation in Lake Chaohu using in-situ stationary measurements with extended integration times at discrete sites. The work provides insight into carbon spatial heterogeneity in a large eutrophic lake and establishes the sensor as a versatile analytical platform with strong potential for environmental monitoring, geochemical degassing assessment, and CCS leakage detection.

    2. Theory and Methodology

    QEPAS is based on the photoacoustic effect. When gas molecules absorb a wavelength-modulated laser beam, periodic photothermal excitation generates acoustic waves within the gas medium. In QEPAS, these acoustic waves are detected with high sensitivity using a quartz tuning fork with a high quality factor (Q). For a weak absorption system in which the wavelength modulation depth is significantly smaller than the absorption linewidth, the magnitude of the second harmonic (2f) signal, denoted S2f and demodulated at the tuning fork resonant frequency f0 using a lock-in amplifier, can be expressed as[37]S=kαCPQfA,where α is the absorption coefficient per unit concentration of the target species; P is the incident laser power; C is the target gas concentration; and Q and f are the quality factor and resonant frequency of the quartz tuning fork, respectively. A is the resonator cross-section area, and k is a constant describing the microphone transfer function and other system parameters.

    This theoretical model highlights three key factors governing QEPAS signal intensity. Laser and Absorption Line: incident laser power P and absorption line strength (accounted for by α) determine energy input.Quartz Tuning Fork Performance: signal amplitude scales with the Q/f0 ratio; high Q and appropriate resonant frequency f0 are crucial for high sensitivity.Gas Concentration: the signal exhibits linear dependence on C, forming the basis of quantitative detection.

    The physical mechanism of water vapor enhancement in this model arises from improved energy transfer efficiency. CO2 molecules excited in the vibrational–rotational mode (specifically the v3 vibrational transition) relax via two mechanisms: slow radiative decay or rapid non-radiative vibrational-to-translational (V–T) energy transfer through collisions. The relaxation rate Γ is defined as follows: Γ=ΓCO2-CO2+ΓCO2-H2O[H2O]+…,where ΓCO2-H2O is the bimolecular V–T energy transfer rate constant for CO2-H2O collisions. Due to the large dipole moment and near-resonant vibrational energy levels of H2O, ΓCO2-H2O is typically one to two orders of magnitude higher than ΓCO2-CO2 or ΓCO2-N2[28,34]. In humid environments, ΓCO2-H2O[H2O] dominates, resulting in a substantial increase in Γ. According to photoacoustic theory, under adiabatic conditions, the generated photoacoustic pressure is proportional to (1−1/γ)×(Γrad/Γ), where Γrad is the radiative relaxation rate, and γ is the adiabatic index (ratio of specific heats) of the gas mixture. Water vapor significantly increases Γ, thereby reducing the Γrad/Γ  ratio. A smaller radiative contribution means a greater proportion of absorbed photon energy is released as localized heat via V–T relaxation rather than lost via radiation. This enhanced thermal conversion strengthens the acoustic wave pressure, leading to an increased effective S2f signal in Eq. (1).

    This research employs an analytical platform based on off-beam quartz-enhanced photoacoustic spectroscopy (OB-QEPAS). The sensing principle derives from the photoacoustic effect. Light from a wavelength-tunable distributed feedback (DFB) diode laser is frequency-modulated at the center of a specific CO2 absorption line near 2.004 µm. As the modulated laser beam passes through the detection region filled with the sample gas, CO2 molecules periodically absorb optical energy and are excited to higher vibrational–rotational energy levels. These excited molecules subsequently collide with water vapor molecules, which serve as highly efficient non-radiative relaxers, rapidly converting vibrational energy into translational kinetic energy. This cyclic thermal conversion induces periodic heating and cooling of the gas, generating acoustic waves. Weak acoustic waves are detected using a custom OB-QEPAS detection module composed of a quartz tuning fork with a high Q-factor (∼10,000 at 32.768 kHz) and a micro-resonator tube positioned parallel to its side. Unlike the conventional on-beam configuration, the laser beam passes beside the tuning fork rather than being precisely aligned through the narrow gap between the prongs. The micro-resonator tube (length: 6 mm; inner diameter: 0.6 mm; outer diameter: 1.2 mm) acoustically focuses and amplifies the generated acoustic wave, which is then efficiently coupled into the tuning fork. Vibration of the fork activates the piezoelectric effect, producing a weak electrical signal proportional to the gas concentration. This signal is extracted and processed at the tuning fork resonant frequency using a high-precision lock-in amplifier[24,25].

    The integrated architecture of the dissolved CO2 sensing system is illustrated in Fig. 1, which presents a schematic diagram of the OB-QEPAS sensor coupled with vacuum membrane degassing. The optical and acoustic core utilizes a DFB diode laser with a center wavelength of 2.004 µm. The laser is from Nanoplus Corporation. The incident laser power at the absorption line (measured after the fiber collimator) is approximately 8.5 mW. To further support the high selectivity of our sensor, we compared the absorption line strengths of CO2, H2O, CH4, and N2O near 2.004 µm using the HITRAN database (Fig. 2). The closest CH4 line in this spectral window has a line strength that is more than three orders of magnitude weaker than that of CO2. Therefore, potential interference from other common dissolved gases (e.g.,  CH4) is negligible, confirming the excellent CO2 selectivity of the sensor. The collimated output beam is directed to pass immediately adjacent to the micro-resonator tube in the OB-QEPAS module. Dissolved gases are extracted using vacuum membrane degassing. The aqueous sample is pumped into the shell side of a hollow fiber membrane unit at a constant flow using a peristaltic pump, while a vacuum pump is connected to the tube-side outlet. Stable negative pressure drives dissolved CO2 and other gases across the membrane wall according to the partial pressure gradient. This method eliminates the need for a purge carrier gas, simplifying the system and preventing contamination from impurities in the carrier gas. The extracted gas mixture flows directly into the downstream QEPAS gas cell for detection. For signal generation and processing, the laser injection current and temperature are precisely controlled using a dedicated laser controller. A sinusoidal waveform at f0/2 is applied to the laser current for wavelength modulation spectroscopy (WMS). To maintain the modulation frequency precisely at half the tuning fork resonance, the optimal modulation frequency was periodically determined by sweeping the frequency and identifying the maximum second harmonic (2f) signal amplitude. The electrical signal from the QEPAS module is demodulated using 2f detection via a lock-in amplifier. To exploit the water vapor enhancement effect and maintain high measurement stability, the complete gas flow path, which spans from the membrane degasser outlet to the QEPAS gas cell, is housed in a precision temperature-controlled enclosure. The device maintains gas temperature within ±0.1°C, ensuring constant, saturated water vapor content in the gas stream. This controlled humidity provides stable and significant enhancement of the photoacoustic signal[30].

    Schematic diagram of the integrated OB-QEPAS dissolved CO2 sensing system with vacuum membrane degassing.

    Figure 1.Schematic diagram of the integrated OB-QEPAS dissolved CO2 sensing system with vacuum membrane degassing.

    Comparison of absorption line strengths of CO2, H2O, CH4, and N2O near 2.004 µm based on the HITRAN database.

    Figure 2.Comparison of absorption line strengths of CO2, H2O, CH4, and N2O near 2.004 µm based on the HITRAN database.

    In QEPAS measurements, the 2f signal often exhibits baseline drift caused by temperature-dependent electronic drift, low-frequency environmental noise, and laser power fluctuations. This drift introduces significant error in concentration quantification, particularly near the detection limit. The measured signal deviates from the ideal form in Eq. (1) and can be expressed as follows: S=kαCPQfA+b(n)+ε(n),where b(n) represents baseline drift and ε(n) denotes random noise.

    A baseline correction algorithm based on endpoint fitting was developed to address this issue. Stable reference regions are selected at the spectral scan extremities (Rstart=1,2,3,4,5, Rend=52,53,54,55,56,57), where the laser wavelength is sufficiently detuned from the absorption line center and gas absorption is negligible. Baseline values are calculated by averaging the signal within these regions: bstart=1|Rstart|∑i∈RstartS2fmeas(i),bend=1|Rend|∑i∈RendS2fmeas(i).

    A linear baseline is constructed as follows: b(n)=bstart−bend−bstartN−1·(n−1).

    The corrected signal is obtained as follows: S2fcorr(n)=S2fmeas(n)−b(n).

    After baseline correction, the signal is optimized to conform to theoretical second harmonic characteristics. Processing includes amplitude normalization [Snorm(n)=S2fcorr(n)/max(|S2fcorr|)], phase alignment to initiate the signal at the positive half-cycle, zero-phase FIR filtering to enhance frequency components, and symmetry adjustment using the symmetry factor α=mean(|Spos|)/mean(|Sneg|) to balance positive and negative half-cycles. Key parameters are extracted for quality evaluation, including peak and valley detection using amplitude thresholds θp and θv, zero-crossing identification, peak-to-peak amplitude [Vpp=max(S)−min(S)], and signal-to-noise ratio [SNR=20 log10(σsignal/σnoise)]. This processing approach ensures consistency with physical principles, maintains computational efficiency O(N), and provides robustness against signal fluctuations, significantly improving measurement accuracy for detecting subtle concentration gradients in environmental monitoring. Figure 3 illustrates the comprehensive workflow of the signal processing pipeline, encompassing the steps from raw signal acquisition to baseline-corrected and optimized output.

    Signal processing workflow for QEPAS baseline correction and optimization.

    Figure 3.Signal processing workflow for QEPAS baseline correction and optimization.

    3. Results and Discussion

    Operating pressure has two competing effects on the 2f signal amplitude. Lower pressure enhances the V–T relaxation rate by increasing the mean free path, but also narrows the absorption line, reducing the overlap with the fixed modulation depth. Higher pressure increases pressure broadening, improving the overlap, but can quench the photoacoustic signal. The observed maximum at 250 Torr represents the optimal trade-off between these two regimes, where the relaxation dynamics and spectral linewidth are best matched to our modulation parameters. As shown in Fig. 4, the 2f signal amplitude exhibited characteristic non-monotonic dependence on pressure. As pressure decreased from ambient levels, reduced intermolecular collision frequency caused vibrational energy relaxation to rely more heavily on resonant energy transfer to specific molecules (e.g.,  H2O), improving photoacoustic conversion efficiency and enhancing the signal. At pressures lower than 100 Torr, substantial spectral line narrowing occurred, resulting in an excessive fixed wavelength modulation depth relative to the absorption linewidth and reducing excitation efficiency. Experimental results identified maximum signal response at 250 Torr. This operating pressure represents an optimal balance between relaxation dynamics and spectral line shape and was therefore adopted as the standard condition for all subsequent measurements.

    Dependence of QEPAS 2f signal amplitude on operating pressure.

    Figure 4.Dependence of QEPAS 2f signal amplitude on operating pressure.

    2f signals used for baseline correction analysis were acquired using a 100 ppm CO2 standard gas under optimized QEPAS operating conditions. System operating pressure was maintained at 250 Torr, and the laser modulation frequency was set to half of the quartz tuning fork resonant frequency (16329.9 Hz). Integration time per data point was 1 s. To ensure measurement stability while preserving the beneficial water vapor enhancement effect, the entire gas flow path was thermally regulated at 28.0°C±0.1°C using a precision temperature control system. A constant gas flow rate of 100 mL/min was maintained. A complete spectral scan across the CO2 absorption line at 2.004 µm generated 57 consecutive data points, providing sufficient spectral resolution for comprehensive baseline characterization and rigorous algorithm validation. Raw 2f signals displayed significant baseline drift, primarily caused by electronic thermal drift, with a magnitude of approximately 15% of the peak signal amplitude. This drift would introduce unacceptable quantification errors for weak signals. The endpoint baseline correction method described in Section 2 was applied to remove the drift. As shown in Fig. 5, the corrected waveform retains the expected harmonic features with drift eliminated, revealing the true harmonic response centered near zero amplitude.

    Comparison of raw and baseline-corrected second harmonic signals.

    Figure 5.Comparison of raw and baseline-corrected second harmonic signals.

    Signal-to-noise ratio (SNR) before and after processing.

    Figure 6.Signal-to-noise ratio (SNR) before and after processing.

    Performance of the full processing pipeline was evaluated using multiple metrics. Baseline drift was reduced by 94.2% relative to initial magnitude. SNR improved by 16.62 dB (Fig. 6). The feature detection algorithm identified three distinct peaks and two valleys with positional accuracy better than 0.1 data point. Zero-crossing determination reached sub-point resolution. Peak-to-peak amplitude variability decreased by 6.2%. The processed signal exhibited strong agreement with the theoretical second harmonic profile, with a correlation coefficient of 0.999 in Fig. 7(a). Under optimized operating conditions, the quantitative calibration curve of the sensor was established using CO2 standard gases with concentrations of 101, 206, 296, 507, 611, 996, and 2000 ppm. Corresponding 2f signal amplitudes are shown in Fig. 7(b). The sensor exhibited an excellent linear response within this range (101–2000 ppm), with a linear regression coefficient R2>0.999. This high degree of linearity confirms quantitative reliability over the full operating range. The system’s detection capability and stability were evaluated using Allan deviation analysis. As shown in Fig. 8, the Allan deviation curve characterizes system noise behavior as a function of averaging time. A detection limit of 4.13 ppm for gaseous CO2 (corresponding to a dissolved CO2 detection limit of 4.13 ppm after degassing calibration) was achieved at a 740 s integration time, indicating that the sensor provides both rapid response and enhanced sensitivity under extended integration. These characteristics support its suitability for long-term, unattended trace gas monitoring.

    (a) Calibration curve showing linear response to CO2 standards. (b) Representative 2f spectral scan.

    Figure 7.(a) Calibration curve showing linear response to CO2 standards. (b) Representative 2f spectral scan.

    Allan deviation analysis demonstrating long-term stability and detection limit of the OB-QEPAS sensor.

    Figure 8.Allan deviation analysis demonstrating long-term stability and detection limit of the OB-QEPAS sensor.

    The synergistic water vapor enhancement effect is fundamental to the sensing strategy. Unlike conventional approaches that reduce water vapor interference, the gas stream is actively maintained in a constant, highly saturated state through precise temperature control. As shown in Fig. 9, the CO2 photoacoustic signal obtained under humidified conditions increased by approximately 1.5× compared with that under dry conditions, consistent with theoretical predictions of water vapor acting as an efficient vibrational energy relaxer. This result confirms that the method successfully converts a typical interferent into a stable signal amplifier.

    Comparison of CO2 photoacoustic signal amplitude under dry and water-vapor-saturated conditions.

    Figure 9.Comparison of CO2 photoacoustic signal amplitude under dry and water-vapor-saturated conditions.

    For the dry and humid comparison, the temperature (28.0°C) and pressure (250 Torr) were kept strictly identical. The enhancement effect was highly reproducible: three repeated measurements yielded an average enhancement factor of 1.5. While we did not perform a continuous humidity titration in this study, previous works[36] have shown that the CO2 QEPAS signal increases monotonically with water vapor concentration. A full characterization of the humidity dependence is planned for future work.

    Following detailed performance assessment, sensor metrics were compared with existing commercial technologies. For long-term in-situ monitoring, gas consumption is a critical parameter, affecting operational cost, maintenance frequency, and deployment flexibility.

    Table 1 summarizes performance indicators of this sensor in comparison with established systems, including TDLAS, CRDS, and electrochemical sensors. The comparison shows that high-sensitivity methods (e.g.,  CRDS) typically lack field robustness and require high operational expenditure, whereas field-deployable systems (e.g.,  electrochemical) often exhibit insufficient sensitivity and long-term drift. The developed sensor is designed specifically to bridge this gap. Although the in-plane QEPAS configuration[32] achieves a detection limit of 1.06 ppm for gaseous CO2, its reliance on precise optical alignment through the narrow prong gap (<500 μm) imposes stringent requirements on beam quality and mechanical stability, limiting its suitability for field deployment. In contrast, the off-beam QEPAS architecture employed in this work inherently tolerates optical misalignment, and the additional baseline correction algorithm further enhances measurement stability under environmental disturbances. The sensor developed in this work demonstrates a well-balanced performance profile. While TDLAS and CRDS achieve lower detection limits, they require high gas consumption and large system volume, and entail higher cost and complexity. In contrast, our sensor achieves a competitive detection limit of 4.13 ppm for dissolved CO2 while offering minimal gas consumption, compact size, low cost, and excellent field robustness—making it an ideal choice for long-term unattended deployment in natural environments.

    TechnologyDetection limit (CO2)Response timeGas consumptionSystem cost and complexityField-deployability
    This work (OB-QEPAS)∼4.13 ppm (dissolved)∼10 sVery low (<10 cm3)aLow–mediumExcellent
    On-beam QEPAS[36]∼18 ppm (gas)∼10 sLow (<10 cm3)LowMedium (alignment sensitive)
    Off-beam QEPAS[38]∼6 ppm (gas)∼10 sLow (<10 cm3)LowGood
    In-plane QEPAS[32]∼1.06 ppm∼10 sLow (<10 cm3)LowMedium
    TDLAS (multi-pass)[39,40]∼0.1–4 ppm∼10 s–1 minMedium–high (150–1500 cm3)Medium–highMedium (vibration sensitive)
    CRDS[41–43]∼0.1–1 ppm∼10–30 sHigh (100–500 cm3)HighPoor (requires vibration isolation)
    Electrochemical[19]∼1–5 ppm∼1–2 minLow (diffusion only)LowGood (prone to drift)

    Table 1. Performance Comparison of CO2 Sensing Technologies

    Extremely low gas consumption is a key advantage, making it ideal for long-term unattended deployment. This combination of robustness, low cost, and very low gas demand provides an optimal balance between performance and field applicability. To validate the operational stability and data acquisition capability of the sensor in real-world environments, in-situ stationary measurements were conducted at five representative sampling sites in the near-surface water of Lake Chaohu. During the measurement campaign, the sensor operated continuously, and all signal acquisition and data processing procedures were completed automatically by the system. No performance degradation or data interruption occurred due to environmental disturbances (e.g.,  boat vibrations, temperature fluctuations). All water samples were collected from a fixed depth of 0.5 m below the surface to ensure consistency in this proof-of-principle field validation. Depth-profile measurements were not performed in this campaign, which is acknowledged as a limitation and will be addressed in future studies.

    2f signals acquired at each site are shown in Fig. 10. The measurement data indicate that, within the observation window of this campaign, the dissolved CO2 concentrations measured at the five sites were of a similar order of magnitude, with an average background concentration of approximately 1058 ppm. This result preliminarily validates the reliable operation of the sensor under field conditions. It must be clearly acknowledged that the experimental design of this measurement campaign has a fundamental methodological limitation: the measurements at the five sites were conducted sequentially at different times throughout a single day, rather than synchronously at all sites at the same moment. Therefore, two possible interpretations exist for the observed similarity in concentrations: 1) it may reflect true spatial homogeneity (i.e., the background concentrations at different locations on the lake surface are indeed similar); or 2) it may simply arise from temporal coincidence (i.e., the concentrations at different sites were actually different, but happened to be at similar levels at their respective sampling times). Given that dissolved CO2 concentrations in shallow lakes can exhibit significant diurnal variations influenced by sunlight, temperature, and biological activity rhythms, the latter possibility cannot be excluded.

    (a) Dissolved CO2 2f amplitude. (b) Dissolved CO2 2f concentration.

    Figure 10.(a) Dissolved CO2 2f amplitude. (b) Dissolved CO2 2f concentration.

    4. Conclusion

    A robust and field-deployable sensor for dissolved CO2 based on off-beam quartz-enhanced photoacoustic spectroscopy (OB-QEPAS) was developed and validated. The core innovation lies in integrating vacuum membrane degassing with a humidity-enhanced OB-QEPAS detection module and a baseline correction-based signal optimization algorithm. The result is a system suitable for long-term operation while maintaining competitive detection sensitivity.

    The main outcomes of this work are as follows.

    Technical Innovation: an active humidity control strategy was implemented to leverage the water vapor enhancement effect, amplifying the photoacoustic signal by approximately 1.5× and improving detection sensitivity. A signal processing pipeline incorporating endpoint baseline correction and optimization enhanced the signal-to-noise ratio by 16.62 dB. Coupled with the rugged OB-QEPAS architecture and carrier-gas-free membrane degassing, the system achieved a detection limit of 4.13 ppm for dissolved CO2 with minimal gas consumption, offering a balanced trade-off among sensitivity, robustness, and field-deployability.

    Analytical Performance: the sensor demonstrated excellent linearity (R2>0.999) and high stability, confirmed by Allan deviation analysis. Low gas consumption, portability, and autonomy make it suitable for long-term deployment in remote or resource-constrained environments.

    Field Validation: multi-site continuous measurements in Lake Chaohu validated the operational stability and data acquisition capability of the sensor in real-world environments. The sensor operated stably under unattended conditions, with preliminary data showing comparable dissolved CO2 concentration levels across the measured sites. This field campaign explicitly identified a methodological limitation—the confounding effect of temporal variability due to non-synchronous sampling—underscoring the need for more rigorous experimental designs (e.g.,  simultaneous sensor deployment, establishment of diurnal baselines) in future studies to elucidate the spatial distribution characteristics of dissolved CO2 in natural waters.

    The innovative coupling of OB-QEPAS robustness, vacuum membrane degassing convenience, and active water vapor enhancement provides a new solution for high-precision in-situ monitoring of dissolved CO2. Further development will continue along technical and scientific lines. Technically, efforts will focus on expanding the platform to multi-component gas sensing via integration of multi-channel degassers and multi-laser systems for simultaneous monitoring of gases such as CH4 and N2O. Additional work will prioritize sensor miniaturization, full integration, and low-power design to enable deployment on autonomous mobile platforms. Scientifically, future studies will employ more rigorous experimental designs to investigate the spatiotemporal distribution characteristics of dissolved gases in aquatic systems and their controlling factors.

    While our sensor uses a standard 32.768 kHz quartz tuning fork, the performance can be further improved by employing custom-designed tuning forks. Recent studies have shown that forks with lower resonance frequencies (e.g.,  10–20 kHz) or optimized geometries can achieve higher Q/f0 ratios, leading to significantly enhanced detection sensitivity[44–47]. Integrating such custom quartz tuning forks into our OB-QEPAS platform represents a promising direction for achieving sub-ppm detection limits for dissolved CO2 and other trace gases.

    References

    [1] N. Kampman, M. Bickle, J. Becker et al. Drilling and sampling of a natural CO2 reservoir: implications for fluid flow and CO2-fluid-rock reactions during CO2 migration through the overburden. Chem. Geol., 369, 51(2014).

    [2] G. Chiodini, C. Cardellini, F. Di Luccio et al. Correlation between tectonic CO2 Earth degassing and seismicity is revealed by a 10-year record in the Apennines, Italy. Sci. Adv., 6, eabc2938(2020).

    [3] P. A. Raymond, J. Hartmann, R. Lauerwald et al. Global carbon dioxide emissions from inland waters. Nature, 503, 355(2013).

    [4] L. J. Tranvik, J. A. Downing, J. B. Cotner et al. Lakes and reservoirs as regulators of carbon cycling and climate. Limnol. Oceanogr., 54, 2298(2009).

    [5] S. Hurwitz, L. E. Clor, R. B. McCleskey et al. Dissolved gases in hydrothermal (phreatic) and geyser eruptions at Yellowstone National Park, USA. Geology, 44, 235(2016).

    [6] Z. Yu, S. C. Singh, C. Hamelin et al. Deep mantle earthquakes linked to CO2 degassing at the Mid-Atlantic Ridge. Nat. Commun., 16, 563(2025).

    [7] L. J. Alcott, C. Walton, N. J. Planavsky et al. Crustal carbonate build-up as a driver for Earth’s oxygenation. Nat. Geosci., 17, 458(2024).

    [8] J. Cao, H. Yang, Z. Kang. Preliminary regional estimation of carbon sink flux by carbonate rock corrosion: a case study of the Pearl River Basin. Chin. Sci. Bull., 56, 3766(2011).

    [9] C. Preston, M. Monea, W. Jazrawi et al. IEA GHG Weyburn CO2 monitoring and storage project. Fuel Process. Technol., 86, 1547(2005).

    [10] M. M. R. Jahangir, P. Johnston, M. I. Khalil et al. Evaluation of headspace equilibration methods for quantifying greenhouse gases in groundwater. J. Environ. Manage., 111, 208(2012).

    [11] J. P. Donval, V. Guyader. Analysis of hydrogen and methane in seawater by ‘headspace’ method: determination at trace level with an automatic headspace sampler. Talanta, 162, 408(2017).

    [12] C. J. C. Van De Ven, I. R. Hearne, M. M. Calvert. Multicomponent mass transfer in dissolved gas analysis: the impacts of headspace pressurization on reliable measurement. ACS EST Water, 5, 639(2025).

    [13] J. Hodgkinson, R. P. Tatam. Optical gas sensing: a review. Meas. Sci. Technol., 24, 012004(2013).

    [14] J. Dai, B. Luo, X. Shen et al. A review of optical gas sensing technology for dissolved gas analysis in transformer oil. Front. Phys., 13, 1547563(2025).

    [15] Z. Li, W. Li, J. Lin et al. Review on performance enhancement methods of tunable diode laser absorption spectroscopy for in situ gas measurement. Appl. Spectrosc. Rev., 60, 555(2025).

    [16] J. Li, B. Guo, X. Sun et al. Current pulse width modulation-cavity ring down spectroscopy for simultaneous measurement of carbon dioxide and methane. Sens. Actuators B Chem., 444, 138457(2025).

    [17] J. W. Pohlman, M. Casso, C. Magen et al. Discrete sample introduction module for quantitative and isotopic analysis of methane and other gases by cavity ring-down spectroscopy. Environ. Sci. Technol., 55, 12066(2021).

    [18] H. C. Ma, Z. M. Yang, H. Y. Cao et al. Triphenylamine-based fluorescent sensors for CO2 response and detection of F− with high sensitivity and selectivity. Dyes Pigments, 140, 131(2017).

    [19] M. Struzik, I. Garbayo, R. Pfenninger et al. A simple and fast electrochemical CO2 sensor based on Li7La3Zr2O12 for environmental monitoring. Adv. Mater., 30, 1804098(2018).

    [20] B. Li, X. Yang, Z. C. Sun et al. Low-cost hydrogen sensor with the transient response of a quartz tuning fork. Sens. Actuators B Chem., 443, 138317(2025).

    [21] C. Sang, C. Zhang, R. Wang et al. Spherical acoustic resonator-based quartz-enhanced photoacoustic-photothermal dual spectroscopy sensing. Anal. Chem., 97, 20536(2025).

    [22] P. Patimisco, G. Scamarcio, F. Tittel et al. Quartz-enhanced photoacoustic spectroscopy: a review. Sensors, 14, 6165(2014).

    [23] Y. Chen, H. Ma, S. Qiao et al. Rapid ppb-level methane detection based on quartz-enhanced photoacoustic spectroscopy. Anal. Chem., 97, 6780(2025).

    [24] S. Torino, L. Conte, M. Iodice et al. PDMS membranes as sensing element in optical sensors for gas detection in water. Sens. Bio-Sens. Res., 16, 74(2017).

    [25] K. Liu, X. Y. Guo, H. M. Yi et al. Off-beam quartz-enhanced photoacoustic spectroscopy. Opt. Lett., 34, 1594(2009).

    [26] P. Breitegger, B. Schweighofer, H. Wegleiter et al. Towards low-cost QEPAS sensors for nitrogen dioxide detection. Photoacoustics, 18, 100169(2020).

    [27] Y. Zhang, H. H. Lv, H. D. Zheng et al. Temperature-compensated calibration for quartz-enhanced photoacoustic spectroscopy by thermoelastic effects. Infrared Phys. Technol., 117, 103816(2021).

    [28] H. Wu, L. Dong, H. Zheng et al. Beat frequency quartz-enhanced photoacoustic spectroscopy for fast and calibration-free continuous trace-gas monitoring. Nat. Commun., 8, 15331(2017).

    [29] M. Mordmüller, M. Köhring, W. Schade et al. An electrically and optically cooperated QEPAS device for highly integrated gas sensors. Appl. Phys. B, 119, 111(2015).

    [30] H. Yi, K. Liu, S. Sun et al. Theoretical analysis of off beam quartz-enhanced photoacoustic spectroscopy sensor. Opt. Commun., 285, 5306(2012).

    [31] Z. Shang, H. Wu, S. Li et al. Elliptical-tube off-beam quartz-enhanced photoacoustic spectroscopy. Appl. Phys. Lett., 120, 171101(2022).

    [32] Y. Xie, Y. X. Zhang, D. H. Zhu et al. Compact QEPAS CO2 sensor system using a quartz tuning fork-embedded and in-plane configuration. Opt. Laser Technol., 159, 109018(2023).

    [33] M. Olivieri, M. Giglio, S. D. Russo et al. Assessment of vibrational-translational relaxation dynamics of methane isotopologues in a wet-nitrogen matrix through QEPAS. Photoacoustics, 31, 100518(2023).

    [34] G. Wysocki, A. Kosterev, F. Tittel. Influence of molecular relaxation dynamics on quartz-enhanced photoacoustic detection of CO2 at λ = 2 µm. Appl. Phys. B, 85, 301(2006).

    [35] A. Kosterev, Y. Bakhirkin, F. Tittel et al. QEPAS methane sensor performance for humidified gases. Appl. Phys. B, 92, 103(2008).

    [36] R. Lewicki, G. Wysocki, A. Kosterev et al. Carbon dioxide and ammonia detection using 2 µm diode laser based quartz-enhanced photoacoustic spectroscopy. Appl. Phys. B, 87, 157(2007).

    [37] A. A. Kosterev, F. K. Tittel, D. V. Serebryakov et al. Applications of quartz tuning forks in spectroscopic gas sensing. Rev. Sci. Instrum., 76, 043105(2005).

    [38] Y. C. Xie, R. F. Wang, Y. Cao et al. Research on detecting CO2 with off-beam quartz-enhanced photoacoustic spectroscopy at 2.004 µm. Spectrosc. Spect. Anal., 40, 2664(2020).

    [39] M. Gu, J. Chen, Y. Zhang et al. Portable TDLAS sensor for online monitoring of CO2 and H2O using a miniaturized multi-pass cell. Sensors, 23, 2072(2023).

    [40] M. Gu, J. Chen, J. Mei et al. Open-path anti-pollution multi-pass cell-based TDLAS sensor for the online measurement of atmospheric H2O and CO2 fluxes. Opt. Express, 30, 43961(2022).

    [41] L. Zhang, Y. Hua, Y. S. Zhou et al. Near-infrared mobile cloud OA-ICOS sensor system for atmospheric carbon dioxide monitoring. Anal. Chem., 97, 4021(2025).

    [42] Z. B. Li, H. L. Ma, Z. S. Cao et al. High-sensitive off-axis integrated cavity output spectroscopy and its measurement of ambient CO2 at 2 µm. Acta Phys. Sin., 65, 053301(2016).

    [43] R. Yu, H. Xia, T. Pang et al. Simultaneous detection of CO2/CH4 based on off-axis integrated cavity output spectroscopy and time-division-multiplexing-based wavelength modulation spectroscopy. Opt. Commun., 545, 129731(2023).

    [44] S. Qiao, X. Liu, Z. Lang et al. Quartz-enhanced laser spectroscopy sensing. Light Sci. Appl., 15, 5(2026).

    [45] R. Wang, S. Qiao, Y. He et al. Highly sensitive laser spectroscopy sensing based on a novel four-prong quartz tuning fork. Opto-Electron. Adv., 8, 240275(2025).

    [46] H. Sun, S. Qiao, Y. He et al. Parts-per-quadrillion level gas molecule detection: CO-LITES sensing. Light Sci. Appl., 14, 180(2025).

    [47] H. Lin, H. Zheng, W. Zhu et al. Multifunctional lithium niobate platform for photodetection and photoacoustic and thermoelastic gas sensing. Nat. Commun., 17, 2296(2026).

    Yingchao Xie, Yuqing Li, Hongjin Liu, Guishi Wang, Gaopeng Lu, Kun Liu, Chunyan Sun, "Water-vapor-enhanced off-beam QEPAS sensor for in-situ monitoring of dissolved CO2," Chin. Opt. Lett. 24, 103001 (2026)
    Download Citation