- Chinese Optics Letters
- Vol. 24, Issue 10, 103001 (2026)
Abstract
1. Introduction
Carbon dioxide () dissolved in aquatic systems constitutes a critical and dynamic reservoir within the global carbon cycle. Accurate quantification is essential for environmental studies and for resolving fundamental geochemical and geophysical processes[1,2]. In natural waters, constraining fluxes from lakes and rivers to the atmosphere is necessary for closing the global carbon budget[3,4]. Dissolved also functions as a geochemical tracer; anomalous degassing in groundwaters and geothermal fluids is often associated with deep-seated processes such as magma migration and fault activity, serving as a potential precursor to seismic events[5,6]. As a primary agent in water–rock interactions, dissolved regulates the dissolution kinetics of carbonate rocks, drives karst formation, and influences the chemical composition of subsurface fluids[7,8]. In the context of carbon capture and storage (CCS), detecting minute increases in dissolved in overlying aquifers is required for early warning of reservoir leakage[9].
Despite the broad scientific relevance of high-precision, high-spatiotemporal-resolution data acquisition, detecting dissolved using existing methods is analytically challenging. The canonical discrete method, headspace equilibrium followed by gas chromatography, is unsuitable for capturing transient signals or fine spatial heterogeneity due to its low temporal resolution and labor-intensive protocol[10,11]. Commercial field sensors are available but often exhibit insufficient sensitivity, long-term drift, or limited selectivity when applied to complex environmental matrices[12,13]. Optical sensing offers a viable route forward. Tunable diode laser absorption spectroscopy (TDLAS) provides high sensitivity but typically uses multi-pass cells susceptible to misalignment under field-induced vibrations, compromising measurement stability and reliability[14,15]. Cavity ring-down spectroscopy (CRDS) achieves exceptionally low detection limits through extended optical path length but is limited by high cost, operational complexity, and susceptibility to environmental perturbations, restricting deployment in extensive field campaigns or remote areas[16,17]. Fluorescence-based sensors[18] and electrochemical systems[19] are highly portable but frequently suffer from signal drift, cross-sensitivity, and inadequate detection limits for resolving subtle geochemical gradients.
Quartz-enhanced photoacoustic spectroscopy (QEPAS) has emerged as a competitive technology for trace gas sensing due to its compact design, inherent immunity to beam misalignment, and high-Q resonant detection[20,21]. A key advantage of QEPAS is its minimal sample volume requirement. The dead volume of the acoustic detection module (tuning fork and micro-resonators) is typically a few cubic centimeters, compared with the hundreds required to purge and stabilize TDLAS multi-pass cells or CRDS cavities, leading to longer system response times and higher carrier gas demands[22,23]. In dissolved gas analysis, efficient and stable gas extraction is essential for optimal optical detection. This study employs a polydimethylsiloxane (PDMS) hollow fiber membrane as the degassing element. This material provides high permeability and selectivity for non-polar gases such as , enabling rapid and continuous separation of dissolved gases from the aqueous to the gas phase[24].
Sign up for Chinese Optics Letters TOC. Get the latest issue of Chinese Optics Letters delivered right to you!Sign up now
The conventional on-beam QEPAS configuration, which requires precise focusing of the laser beam through the narrow gap () between the tuning fork prongs, achieves high sensitivity in laboratory settings. This precise optical alignment increases vulnerability to performance degradation caused by temperature fluctuations and mechanical vibrations during field deployment, limiting instrumental ruggedness and long-term stability[25]. The narrow resonance profile of the high-Q quartz tuning fork, while beneficial for noise rejection, also makes signal amplitude highly sensitive to drifts in the laser wavelength modulation frequency. If uncorrected, such drifts can cause significant signal loss and measurement instability over time, particularly under variable environmental conditions in field applications[26,27]. Active frequency tracking techniques, such as using a phase-locked loop (PLL) to lock the modulation frequency to the fork’s resonant frequency in real time or periodically sweeping the modulation frequency to locate the maximum response, are commonly implemented to maintain optimal and stable operation[28,29]. The off-beam QEPAS (OB-QEPAS) configuration was developed to overcome the optical alignment limitation[30]. By directing the laser beam adjacent to the tuning fork, alignment tolerances are substantially relaxed and mechanical robustness is improved, making the sensor architecture suitable for long-term deployment in demanding field environments, including mobile platforms and seismically active regions[31,32].
The influence of water vapor on QEPAS performance depends on the spectral line and can be managed strategically. Spectral line selection using the HITRAN database[33,34] confirms that direct absorption interference from is negligible at the 2.004 µm absorption line used in this study. Water vapor molecules act as efficient vibrational energy transfer partners for excited , amplifying the photoacoustic signal through resonant vibrational-to-translational (V–T) energy relaxation[33,34]. Instead of treating water vapor as an interferent requiring removal, which reduces signal strength, this work harnesses the “water vapor enhancement effect” as a stable signal amplification mechanism. Precise thermal control of the gas stream exiting the membrane equilibrator[35] is used to actively maintain consistently high water vapor saturation. This approach stabilizes the measurement matrix and provides substantial and consistent enhancement of the photoacoustic signal, improving in-situ detection sensitivity[36].
Combining the rugged OB-QEPAS architecture with efficient sample pretreatment via PDMS membrane degassing, controlled humidity-based signal enhancement, advanced signal processing for drift and noise suppression, and robust frequency stabilization, this study develops a portable analyzer for dissolved . Field capability was validated through a field validation in Lake Chaohu using in-situ stationary measurements with extended integration times at discrete sites. The work provides insight into carbon spatial heterogeneity in a large eutrophic lake and establishes the sensor as a versatile analytical platform with strong potential for environmental monitoring, geochemical degassing assessment, and CCS leakage detection.
2. Theory and Methodology
QEPAS is based on the photoacoustic effect. When gas molecules absorb a wavelength-modulated laser beam, periodic photothermal excitation generates acoustic waves within the gas medium. In QEPAS, these acoustic waves are detected with high sensitivity using a quartz tuning fork with a high quality factor (Q). For a weak absorption system in which the wavelength modulation depth is significantly smaller than the absorption linewidth, the magnitude of the second harmonic () signal, denoted and demodulated at the tuning fork resonant frequency using a lock-in amplifier, can be expressed as[37]
This theoretical model highlights three key factors governing QEPAS signal intensity.
The physical mechanism of water vapor enhancement in this model arises from improved energy transfer efficiency. molecules excited in the vibrational–rotational mode (specifically the vibrational transition) relax via two mechanisms: slow radiative decay or rapid non-radiative vibrational-to-translational (V–T) energy transfer through collisions. The relaxation rate is defined as follows:
This research employs an analytical platform based on off-beam quartz-enhanced photoacoustic spectroscopy (OB-QEPAS). The sensing principle derives from the photoacoustic effect. Light from a wavelength-tunable distributed feedback (DFB) diode laser is frequency-modulated at the center of a specific absorption line near 2.004 µm. As the modulated laser beam passes through the detection region filled with the sample gas, molecules periodically absorb optical energy and are excited to higher vibrational–rotational energy levels. These excited molecules subsequently collide with water vapor molecules, which serve as highly efficient non-radiative relaxers, rapidly converting vibrational energy into translational kinetic energy. This cyclic thermal conversion induces periodic heating and cooling of the gas, generating acoustic waves. Weak acoustic waves are detected using a custom OB-QEPAS detection module composed of a quartz tuning fork with a high Q-factor ( at 32.768 kHz) and a micro-resonator tube positioned parallel to its side. Unlike the conventional on-beam configuration, the laser beam passes beside the tuning fork rather than being precisely aligned through the narrow gap between the prongs. The micro-resonator tube (length: 6 mm; inner diameter: 0.6 mm; outer diameter: 1.2 mm) acoustically focuses and amplifies the generated acoustic wave, which is then efficiently coupled into the tuning fork. Vibration of the fork activates the piezoelectric effect, producing a weak electrical signal proportional to the gas concentration. This signal is extracted and processed at the tuning fork resonant frequency using a high-precision lock-in amplifier[24,25].
The integrated architecture of the dissolved sensing system is illustrated in Fig. 1, which presents a schematic diagram of the OB-QEPAS sensor coupled with vacuum membrane degassing. The optical and acoustic core utilizes a DFB diode laser with a center wavelength of 2.004 µm. The laser is from Nanoplus Corporation. The incident laser power at the absorption line (measured after the fiber collimator) is approximately 8.5 mW. To further support the high selectivity of our sensor, we compared the absorption line strengths of , , , and near 2.004 µm using the HITRAN database (Fig. 2). The closest line in this spectral window has a line strength that is more than three orders of magnitude weaker than that of . Therefore, potential interference from other common dissolved gases (e.g., ) is negligible, confirming the excellent selectivity of the sensor. The collimated output beam is directed to pass immediately adjacent to the micro-resonator tube in the OB-QEPAS module. Dissolved gases are extracted using vacuum membrane degassing. The aqueous sample is pumped into the shell side of a hollow fiber membrane unit at a constant flow using a peristaltic pump, while a vacuum pump is connected to the tube-side outlet. Stable negative pressure drives dissolved and other gases across the membrane wall according to the partial pressure gradient. This method eliminates the need for a purge carrier gas, simplifying the system and preventing contamination from impurities in the carrier gas. The extracted gas mixture flows directly into the downstream QEPAS gas cell for detection. For signal generation and processing, the laser injection current and temperature are precisely controlled using a dedicated laser controller. A sinusoidal waveform at is applied to the laser current for wavelength modulation spectroscopy (WMS). To maintain the modulation frequency precisely at half the tuning fork resonance, the optimal modulation frequency was periodically determined by sweeping the frequency and identifying the maximum second harmonic () signal amplitude. The electrical signal from the QEPAS module is demodulated using detection via a lock-in amplifier. To exploit the water vapor enhancement effect and maintain high measurement stability, the complete gas flow path, which spans from the membrane degasser outlet to the QEPAS gas cell, is housed in a precision temperature-controlled enclosure. The device maintains gas temperature within , ensuring constant, saturated water vapor content in the gas stream. This controlled humidity provides stable and significant enhancement of the photoacoustic signal[30].
![]()
Figure 1.Schematic diagram of the integrated OB-QEPAS dissolved CO2 sensing system with vacuum membrane degassing.
![]()
Figure 2.Comparison of absorption line strengths of CO2, H2O, CH4, and N2O near 2.004 µm based on the HITRAN database.
In QEPAS measurements, the signal often exhibits baseline drift caused by temperature-dependent electronic drift, low-frequency environmental noise, and laser power fluctuations. This drift introduces significant error in concentration quantification, particularly near the detection limit. The measured signal deviates from the ideal form in Eq. (1) and can be expressed as follows:
A baseline correction algorithm based on endpoint fitting was developed to address this issue. Stable reference regions are selected at the spectral scan extremities (, ), where the laser wavelength is sufficiently detuned from the absorption line center and gas absorption is negligible. Baseline values are calculated by averaging the signal within these regions:
A linear baseline is constructed as follows:
The corrected signal is obtained as follows:
After baseline correction, the signal is optimized to conform to theoretical second harmonic characteristics. Processing includes amplitude normalization , phase alignment to initiate the signal at the positive half-cycle, zero-phase FIR filtering to enhance frequency components, and symmetry adjustment using the symmetry factor to balance positive and negative half-cycles. Key parameters are extracted for quality evaluation, including peak and valley detection using amplitude thresholds and , zero-crossing identification, peak-to-peak amplitude , and signal-to-noise ratio . This processing approach ensures consistency with physical principles, maintains computational efficiency , and provides robustness against signal fluctuations, significantly improving measurement accuracy for detecting subtle concentration gradients in environmental monitoring. Figure 3 illustrates the comprehensive workflow of the signal processing pipeline, encompassing the steps from raw signal acquisition to baseline-corrected and optimized output.
![]()
Figure 3.Signal processing workflow for QEPAS baseline correction and optimization.
3. Results and Discussion
Operating pressure has two competing effects on the signal amplitude. Lower pressure enhances the V–T relaxation rate by increasing the mean free path, but also narrows the absorption line, reducing the overlap with the fixed modulation depth. Higher pressure increases pressure broadening, improving the overlap, but can quench the photoacoustic signal. The observed maximum at 250 Torr represents the optimal trade-off between these two regimes, where the relaxation dynamics and spectral linewidth are best matched to our modulation parameters. As shown in Fig. 4, the signal amplitude exhibited characteristic non-monotonic dependence on pressure. As pressure decreased from ambient levels, reduced intermolecular collision frequency caused vibrational energy relaxation to rely more heavily on resonant energy transfer to specific molecules (e.g., ), improving photoacoustic conversion efficiency and enhancing the signal. At pressures lower than 100 Torr, substantial spectral line narrowing occurred, resulting in an excessive fixed wavelength modulation depth relative to the absorption linewidth and reducing excitation efficiency. Experimental results identified maximum signal response at 250 Torr. This operating pressure represents an optimal balance between relaxation dynamics and spectral line shape and was therefore adopted as the standard condition for all subsequent measurements.
![]()
Figure 4.Dependence of QEPAS 2f signal amplitude on operating pressure.
signals used for baseline correction analysis were acquired using a 100 ppm standard gas under optimized QEPAS operating conditions. System operating pressure was maintained at 250 Torr, and the laser modulation frequency was set to half of the quartz tuning fork resonant frequency (16329.9 Hz). Integration time per data point was 1 s. To ensure measurement stability while preserving the beneficial water vapor enhancement effect, the entire gas flow path was thermally regulated at using a precision temperature control system. A constant gas flow rate of 100 mL/min was maintained. A complete spectral scan across the absorption line at 2.004 µm generated 57 consecutive data points, providing sufficient spectral resolution for comprehensive baseline characterization and rigorous algorithm validation. Raw signals displayed significant baseline drift, primarily caused by electronic thermal drift, with a magnitude of approximately 15% of the peak signal amplitude. This drift would introduce unacceptable quantification errors for weak signals. The endpoint baseline correction method described in Section 2 was applied to remove the drift. As shown in Fig. 5, the corrected waveform retains the expected harmonic features with drift eliminated, revealing the true harmonic response centered near zero amplitude.
![]()
Figure 5.Comparison of raw and baseline-corrected second harmonic signals.
![]()
Figure 6.Signal-to-noise ratio (SNR) before and after processing.
Performance of the full processing pipeline was evaluated using multiple metrics. Baseline drift was reduced by 94.2% relative to initial magnitude. SNR improved by 16.62 dB (Fig. 6). The feature detection algorithm identified three distinct peaks and two valleys with positional accuracy better than 0.1 data point. Zero-crossing determination reached sub-point resolution. Peak-to-peak amplitude variability decreased by 6.2%. The processed signal exhibited strong agreement with the theoretical second harmonic profile, with a correlation coefficient of 0.999 in Fig. 7(a). Under optimized operating conditions, the quantitative calibration curve of the sensor was established using standard gases with concentrations of 101, 206, 296, 507, 611, 996, and 2000 ppm. Corresponding signal amplitudes are shown in Fig. 7(b). The sensor exhibited an excellent linear response within this range (101–2000 ppm), with a linear regression coefficient . This high degree of linearity confirms quantitative reliability over the full operating range. The system’s detection capability and stability were evaluated using Allan deviation analysis. As shown in Fig. 8, the Allan deviation curve characterizes system noise behavior as a function of averaging time. A detection limit of 4.13 ppm for gaseous (corresponding to a dissolved detection limit of 4.13 ppm after degassing calibration) was achieved at a 740 s integration time, indicating that the sensor provides both rapid response and enhanced sensitivity under extended integration. These characteristics support its suitability for long-term, unattended trace gas monitoring.
![]()
Figure 7.(a) Calibration curve showing linear response to CO2 standards. (b) Representative 2f spectral scan.
![]()
Figure 8.Allan deviation analysis demonstrating long-term stability and detection limit of the OB-QEPAS sensor.
The synergistic water vapor enhancement effect is fundamental to the sensing strategy. Unlike conventional approaches that reduce water vapor interference, the gas stream is actively maintained in a constant, highly saturated state through precise temperature control. As shown in Fig. 9, the photoacoustic signal obtained under humidified conditions increased by approximately compared with that under dry conditions, consistent with theoretical predictions of water vapor acting as an efficient vibrational energy relaxer. This result confirms that the method successfully converts a typical interferent into a stable signal amplifier.
![]()
Figure 9.Comparison of CO2 photoacoustic signal amplitude under dry and water-vapor-saturated conditions.
For the dry and humid comparison, the temperature (28.0°C) and pressure (250 Torr) were kept strictly identical. The enhancement effect was highly reproducible: three repeated measurements yielded an average enhancement factor of 1.5. While we did not perform a continuous humidity titration in this study, previous works[36] have shown that the QEPAS signal increases monotonically with water vapor concentration. A full characterization of the humidity dependence is planned for future work.
Following detailed performance assessment, sensor metrics were compared with existing commercial technologies. For long-term in-situ monitoring, gas consumption is a critical parameter, affecting operational cost, maintenance frequency, and deployment flexibility.
Table 1 summarizes performance indicators of this sensor in comparison with established systems, including TDLAS, CRDS, and electrochemical sensors. The comparison shows that high-sensitivity methods (e.g., CRDS) typically lack field robustness and require high operational expenditure, whereas field-deployable systems (e.g., electrochemical) often exhibit insufficient sensitivity and long-term drift. The developed sensor is designed specifically to bridge this gap. Although the in-plane QEPAS configuration[32] achieves a detection limit of 1.06 ppm for gaseous , its reliance on precise optical alignment through the narrow prong gap () imposes stringent requirements on beam quality and mechanical stability, limiting its suitability for field deployment. In contrast, the off-beam QEPAS architecture employed in this work inherently tolerates optical misalignment, and the additional baseline correction algorithm further enhances measurement stability under environmental disturbances. The sensor developed in this work demonstrates a well-balanced performance profile. While TDLAS and CRDS achieve lower detection limits, they require high gas consumption and large system volume, and entail higher cost and complexity. In contrast, our sensor achieves a competitive detection limit of 4.13 ppm for dissolved while offering minimal gas consumption, compact size, low cost, and excellent field robustness—making it an ideal choice for long-term unattended deployment in natural environments.
| Technology | Detection limit (CO2) | Response time | Gas consumption | System cost and complexity | Field-deployability |
|---|---|---|---|---|---|
| This work (OB-QEPAS) | ∼4.13 ppm (dissolved) | ∼10 s | Very low (<10 cm3)a | Low–medium | Excellent ![]() |
| On-beam QEPAS[36] | ∼18 ppm (gas) | ∼10 s | Low (<10 cm3) | Low | Medium (alignment sensitive) |
| Off-beam QEPAS[38] | ∼6 ppm (gas) | ∼10 s | Low (<10 cm3) | Low | Good |
| In-plane QEPAS[32] | ∼1.06 ppm | ∼10 s | Low (<10 cm3) | Low | Medium |
| TDLAS (multi-pass)[39,40] | ∼0.1–4 ppm | ∼10 s–1 min | Medium–high (150–1500 cm3) | Medium–high | Medium (vibration sensitive) |
| CRDS[41–43] | ∼0.1–1 ppm | ∼10–30 s | High (100–500 cm3) | High | Poor (requires vibration isolation) |
| Electrochemical[19] | ∼1–5 ppm | ∼1–2 min | Low (diffusion only) | Low | Good (prone to drift) |
Table 1. Performance Comparison of CO2 Sensing Technologies
Extremely low gas consumption is a key advantage, making it ideal for long-term unattended deployment. This combination of robustness, low cost, and very low gas demand provides an optimal balance between performance and field applicability. To validate the operational stability and data acquisition capability of the sensor in real-world environments, in-situ stationary measurements were conducted at five representative sampling sites in the near-surface water of Lake Chaohu. During the measurement campaign, the sensor operated continuously, and all signal acquisition and data processing procedures were completed automatically by the system. No performance degradation or data interruption occurred due to environmental disturbances (e.g., boat vibrations, temperature fluctuations). All water samples were collected from a fixed depth of 0.5 m below the surface to ensure consistency in this proof-of-principle field validation. Depth-profile measurements were not performed in this campaign, which is acknowledged as a limitation and will be addressed in future studies.
signals acquired at each site are shown in Fig. 10. The measurement data indicate that, within the observation window of this campaign, the dissolved concentrations measured at the five sites were of a similar order of magnitude, with an average background concentration of approximately 1058 ppm. This result preliminarily validates the reliable operation of the sensor under field conditions. It must be clearly acknowledged that the experimental design of this measurement campaign has a fundamental methodological limitation: the measurements at the five sites were conducted sequentially at different times throughout a single day, rather than synchronously at all sites at the same moment. Therefore, two possible interpretations exist for the observed similarity in concentrations: 1) it may reflect true spatial homogeneity (i.e., the background concentrations at different locations on the lake surface are indeed similar); or 2) it may simply arise from temporal coincidence (i.e., the concentrations at different sites were actually different, but happened to be at similar levels at their respective sampling times). Given that dissolved concentrations in shallow lakes can exhibit significant diurnal variations influenced by sunlight, temperature, and biological activity rhythms, the latter possibility cannot be excluded.
![]()
Figure 10.(a) Dissolved CO2 2f amplitude. (b) Dissolved CO2 2f concentration.
4. Conclusion
A robust and field-deployable sensor for dissolved based on off-beam quartz-enhanced photoacoustic spectroscopy (OB-QEPAS) was developed and validated. The core innovation lies in integrating vacuum membrane degassing with a humidity-enhanced OB-QEPAS detection module and a baseline correction-based signal optimization algorithm. The result is a system suitable for long-term operation while maintaining competitive detection sensitivity.
The main outcomes of this work are as follows.
Technical Innovation: an active humidity control strategy was implemented to leverage the water vapor enhancement effect, amplifying the photoacoustic signal by approximately and improving detection sensitivity. A signal processing pipeline incorporating endpoint baseline correction and optimization enhanced the signal-to-noise ratio by 16.62 dB. Coupled with the rugged OB-QEPAS architecture and carrier-gas-free membrane degassing, the system achieved a detection limit of 4.13 ppm for dissolved with minimal gas consumption, offering a balanced trade-off among sensitivity, robustness, and field-deployability.
Analytical Performance: the sensor demonstrated excellent linearity () and high stability, confirmed by Allan deviation analysis. Low gas consumption, portability, and autonomy make it suitable for long-term deployment in remote or resource-constrained environments.
Field Validation: multi-site continuous measurements in Lake Chaohu validated the operational stability and data acquisition capability of the sensor in real-world environments. The sensor operated stably under unattended conditions, with preliminary data showing comparable dissolved concentration levels across the measured sites. This field campaign explicitly identified a methodological limitation—the confounding effect of temporal variability due to non-synchronous sampling—underscoring the need for more rigorous experimental designs (e.g., simultaneous sensor deployment, establishment of diurnal baselines) in future studies to elucidate the spatial distribution characteristics of dissolved in natural waters.
The innovative coupling of OB-QEPAS robustness, vacuum membrane degassing convenience, and active water vapor enhancement provides a new solution for high-precision in-situ monitoring of dissolved . Further development will continue along technical and scientific lines. Technically, efforts will focus on expanding the platform to multi-component gas sensing via integration of multi-channel degassers and multi-laser systems for simultaneous monitoring of gases such as and . Additional work will prioritize sensor miniaturization, full integration, and low-power design to enable deployment on autonomous mobile platforms. Scientifically, future studies will employ more rigorous experimental designs to investigate the spatiotemporal distribution characteristics of dissolved gases in aquatic systems and their controlling factors.
While our sensor uses a standard 32.768 kHz quartz tuning fork, the performance can be further improved by employing custom-designed tuning forks. Recent studies have shown that forks with lower resonance frequencies (e.g., 10–20 kHz) or optimized geometries can achieve higher ratios, leading to significantly enhanced detection sensitivity[44–47]. Integrating such custom quartz tuning forks into our OB-QEPAS platform represents a promising direction for achieving sub-ppm detection limits for dissolved and other trace gases.
References
[38] Y. C. Xie, R. F. Wang, Y. Cao et al. Research on detecting CO2 with off-beam quartz-enhanced photoacoustic spectroscopy at 2.004 µm. Spectrosc. Spect. Anal., 40, 2664(2020).
Set citation alerts for the article